The JINLING 820 Anesthesia Machine operates as a pneumatically driven and electronically controlled system, supporting closed, semi-closed, and semi-open breathing circuits. Housed in a durable, non-hazardous ABS and metal frame, the machine is both environmentally friendly and robust. It features an automated startup self-check and sensor self-calibration to ensure consistent, fault-free performance from the moment it is powered on.
Equipped with a clear 5.7-inch TFT display, this Anesthesia Machine provides real-time visualization of Pressure-Time and Frequency-Time waveforms, which can be toggled instantly with a single button press. The intuitive interface allows anesthesiologists to monitor ventilation parameters efficiently, enhancing situational awareness and procedural accuracy.
The JINLING 820 Anesthesia Machine supports a range of ventilation modes to accommodate diverse clinical needs:
IPPV (VCV)
SIPPV (VCV)
IMV
SIMV
MANUAL
Key ventilation parameters include:
Tidal Volume: 0–1500 mL
Respiratory Frequency: 0–100 bpm
SIMV Frequency: 1–20 bpm
I:E Ratio: 8:1 to 1:10
PEEP Range: 0–2.0 kPa
Inspiratory Pressure Trigger: –1.0 to +2.0 kPa
Inspiratory Plateau: 0–1 second
SIGH Function: One deep breath per 70–120 controlled respirations
A high-precision four-tube flowmeter ensures accurate delivery of O₂ and N₂O, with flow ranges of 0.1–10 L/min for each gas. The system includes an interlocking oxygen-nitrous oxide supply mechanism that guarantees oxygen concentration never falls below 21%. An electronic pressure differential flow sensor and imported respiratory control valve contribute to stable and responsive gas delivery.
The Anesthesia Machine is equipped with a single R-A1 type vaporizer slot, compatible with Enflurane, Isoflurane, and Sevoflurane (Halothane optional). The vaporizer features self-compensating functionality for flow, temperature, and pressure, ensuring consistent anesthetic agent delivery.
Patient safety is reinforced through comprehensive audible and visual alarms for:
Tidal volume
Airway pressure
Asphyxia
Power failure
Oxygen failure
An alarm query function allows quick review of alerts. In case of AC power failure, the Anesthesia Machine automatically switches to backup power mode, ensuring continuous operation without interruption.
The integrated metallic modular respiratory circuit is designed for durability and ease of maintenance, with reliable sealing even under high-temperature disinfection. The machine also includes an advanced rotating soda lime (Na-lime) tank made from imported PPSU material, resistant to temperatures up to 134°C, and allows for quick replacement.
The JINLING 820 Anesthesia Machine can be enhanced with an optional multi-parameter patient monitor, enabling it to function as a comprehensive anesthesia workstation tailored to specific clinical needs.
| Category | Specification | Details |
|---|---|---|
| General | Model Name | JINLING 820 Anesthesia Machine |
| Operating Principle | Pneumatically driven & electronically controlled | |
| Breathing Systems Supported | Closed, semi-closed, semi-open | |
| Frame Material | ABS and Metal | |
| Environmental Design | Non-hazardous, environmentally friendly | |
| Startup Self-Check | Yes, with automatic sensor calibration | |
| Backup Power Support | Yes, automatic switchover during AC power failure | |
| Display & Interface | Display Type | 5.7-inch TFT display |
| Waveforms Displayed | Pressure-Time, Frequency-Time | |
| Waveform Control | Instant cycling with button press | |
| Ventilation | Modes Available | IPPV (VCV), SIPPV (VCV), IMV, SIMV, MANUAL |
| Tidal Volume Range | 0–1500 mL | |
| Respiratory Frequency Range | 0–100 bpm | |
| SIMV Frequency Range | 1–20 bpm | |
| I:E Ratio Range | 8:1 to 1:10 | |
| Inspiratory Plateau Time | 0–1 second | |
| PEEP Range | 0–2.0 kPa | |
| SIGH Function | One deep breath per 70–120 controlled respirations (inspiration time ≥1.5× set value) | |
| Inspiratory Pressure Trigger | –1.0 to +2.0 kPa | |
| Gas Delivery | Flowmeter Type | High-precision four-glass-tube flowmeter |
| O₂ Flow Range | 0.1–10 L/min | |
| N₂O Flow Range | 0.1–10 L/min | |
| O₂ Safety Mechanism | Interlocking O₂/N₂O supply; O₂ concentration never below 21% | |
| Vaporizer | Vaporizer Slot Type | Single R-A1 type slot |
| Compatible Agents | Enflurane, Isoflurane, Sevoflurane (Halothane optional) | |
| Concentration Range | 0–5% | |
| Compensation Features | Automatic flow, temperature, and pressure compensation | |
| Sensors & Valves | Flow Sensor | Electronic pressure differential flow sensor |
| Respiratory Control Valve | Imported, for stable and responsive performance | |
| Breathing Circuit | Circuit Type | Integrated metallic modular respiratory circuit |
| High-Temperature Sealing | Yes, suitable for high-temperature disinfection | |
| Maintenance | Easy to clean and disinfect | |
| CO₂ Absorption | Soda Lime Canister | Advanced international rotating Na-lime tank |
| Material | Imported PPSU | |
| Temperature Resistance | Up to 134°C | |
| Replacement | Designed for easy handling | |
| Alarms & Safety | Alarm Types | Audible and visual alarms |
| Alarm Parameters | Tidal volume, Airway pressure, Asphyxia, Power failure, Oxygen failure | |
| Alarm Query Function | Yes | |
| Optional Features | Multi-parameter Patient Monitor | Optional |
| Anesthesia Workstation Capability | Yes, when combined with optional monitor |